Department of Chemistry
SOPs
Standard Operating Procedures (SOPs) for Management of Departmental Resources
1. Laboratory and Chemical Safety
Labs:
· Each lab (UG and PG practical labs, research, and instrumental labs) will have a safety manual and the students are required to use personal protective equipment before entering the labs.
· Labs will have local exhaust and hoods for proper ventilation and all chemical containers must be properly labeled and arranged alphabetically.
· Potentially hazardous chemical processes or procedures will be carried out by following their specific SOPs, and must not be left unattended by the lab technician.
· Storing food in the lab refrigerator is strictly prohibited.
Chemicals:
· Chemicals, particularly hazardous and high risk must be safely used according to the laboratory-specific SOPs and chemical safety guideline.
· Students must be encouraged to use online resources such as ‘’Laboratory Chemical Safety Summaries (LCSS)’’ published on PubChem and ‘’Prudent Practices in the Laboratory’’ published by the National Research Council for handling and management of chemical hazards.
Equipment:
· Sophisticated instruments and equipment that involve high hazard operations (EPR, PXRD, UV-Vis spectrometry, FTIR, AAS, ITC, and GCMS) must be handled with their special handling procedures for their safe and proper use and be regularly monitored by engineers to maintain their quality.
2. Storage of Chemicals
· Chemicals must be stocked in separate chemical storerooms.
· Each chemical will be kept in defined storage place alphabetically, and toxic and hazardous chemicals will be separately stored in a well-identified area.
· Photosensitive chemicals will be kept in a dark place.
· Chemical and stationary inventory will be updated regularly and accurately.
3. Seminar Library
· The library will purchase the latest books, e-books, journals, e-journals, and newspapers to improve the knowledge and critical thinking of the students, researchers, and teachers.
· Faculty and students after their appointment or enrolment will fill library form with their details to become the member of the library and submit it to library in-charge, and only then library cards will be issued.
· The library will use its own guideline for the borrowing of books and journals.
· All member will follow the general guidelines and instructions issued by the library in-charge for the proper functioning of the library.
SOP For Cell Culture
Lab
1.
Access to the laboratory is limited to facility users only.
2.
Remove all reagents and supplies from the Biosafety Cabinet
upon completion of the experiment, or if there is a long delay in time between
experimental steps.
3.
Users MUST follow all safety and cleanliness guidelines:
a.
Label all reagents and samples stored in common areas
(not-labeled samples will be thrown away during weekly lab clean-up)
b.
Clean the internal surface of the Biosafety Cabinet with 70%
Ethanol/water solution. Bottles with the mixed solutions are available. Please
refill bottles (ethanol and distilled water are available in the lab) upon
usage.
c.
All waist must be properly placed into the biosafety waste
container (containers are provided and will be cleaned weekly).
d.
If User generates a lot of waste and fills out the container,
User should remove the full plastic back from the container, seal it and place
a new bag into the waste container. Bags are located at the bottom of the waste
container.
4.
Persons must wash their hands after working with potentially
hazardous materials and before leaving the laboratory.
5.
Sharps, such as needles, scalpels, pipettes, and broken
glassware must be placed in the appropriate designated containers.
6.
Decontaminate work surfaces after completion of work and
after any spill or splash of potentially infectious material with appropriate
disinfectant.
7.
Decontaminate all cultures, stocks, and other potentially
infectious materials before disposal with bleach or autoclaving.
8.
All procedures involving the manipulation of cells, tissues
or other potentially infectious materials will be performed in a certified
biological safety cabinet and in a manner that minimizes splashing, spraying or
aerosolization.
9.
Before leaving the laboratory, all personal protective
equipment shall be removed and placed in an appropriately designated place for
storage, washing, decontamination or disposal.
10. Used needles and
other contaminated sharps must be discarded immediately after use into a sharps
disposal container. Needles and other sharps shall not be sheared, bent,
broken, recapped or re-sheathed unless required by a specific procedure and it
can be demonstrated that there is no other feasible alternative.
11. Eating,
drinking, smoking, applying cosmetics or lip balm or handling contact lenses is
prohibited in laboratories.
12. Food and Drink
shall not be eaten or stored in the Laboratory.
13. Mechanical
devices are used for all pipetting and suctioning procedures (no mouth
pipetting).
14. Gloves made from
nitrile or another appropriate material will be worn when working with cell and
tissue cultures and other potentially infectious materials.
15. Never touch your
eyes, nose, mouth or your face while wearing gloves.
16. To prevent the
spread of contamination gloves must be removed before touching “clean” surfaces
such as door knobs, computer keyboards, books, telephones/cell phones, etc.
Similarly, gloves used for other laboratory activities should be removed and
new clean pair put on before beginning work with cell and tissue cultures.
17. Do not wash or
reuse disposable gloves.
18. Before starting
work in the BSC, review all procedures that will be used; identify the
necessary equipment and materials that will be needed and develop a plan for
safe and efficient work.
19. If the cabinet
is not running, turn on the blower and fluorescent lights and turn off the UV
light if it is on.
20. Verify that the
BSC is operating correctly:
a.
Check the instrument display/gauges for operational status.
b.
Check the intake and exhaust grills for obstructions.
c.
Check that the sash is in the appropriate position.
d.
Check for the inward flow of air at the face of the BSC by
holding a tissue near the bottom edge of the sash
21. Wipe down the
interior surfaces of the cabinet with an appropriate disinfectant such as 70% ethanol
22. Load the cabinet
with materials that will be needed for the procedure, wiping their surfaces
with 70% ethanol to minimize the introduction of contaminants into the BSC.
Position the materials near the back of the hood and organize them in a manner
that will allow for the separation of clean and contaminated items during your
work in the hood. Only materials needed for immediate work should be place in
the cabinet. Extra supplies (gloves, culture flasks/plates should be stored
outside the cabinet).
23. Before beginning
your work, allow the hood to run for a minimum of 5 minutes to purge any
airborne contaminates from the work area.
24. Avoid rapid,
sweeping movements of the arms into or out of the cabinet. Move items into our
out of the cabinet slowly and perpendicular to the face of the cabinet to
minimize disturbance to the protective curtain of air.
25. Discard all
waste materials generated by your work into appropriate containers inside the
BSC. Close or cover all open containers.
26. Allow the
cabinet to run for 3 to 5 minutes with no activity.
27. Disinfect the
surfaces of all materials, equipment and containers that will be removed from
the BSC, to minimize subsequent contamination in the laboratory.
28. Remove
contaminated gloves and dispose of them appropriately
29. After putting on
a clean pair of gloves, remove all materials for the BSC.
30. Wipe down all
interior surfaces of the BSC with an appropriate disinfectant.
31. If the BSC is
not scheduled for subsequent use, turn off the fluorescent light and cabinet
blower. BSCs are designed for 24 hour operation, but in the interest of energy
conservation it should be shut down when it will not be used for an extended
period of time.
32. Work surfaces
and equipment shall be cleaned and decontaminated on a regular basis with a
freshly made solution of household bleach diluted between 1:10 to 1:100 or
another suitable cleaner/disinfectant.
a.
After completion of procedures involving potentially
infectious materials.
b.
Immediately after a spill of potentially infectious material
and/or when surfaces or equipment become overtly contaminated
Perkin
Elmer Lamda 25
UV/VIS
Spectrometer SOP
The usage of the
machine is NOT allowed without a proper training by the operator of the
facility.
Ø Handling the
instrument
1.
Open the sample compartment cover.
2.
Make sure that beam paths are free,i.e no
objects project into beam paths, no samples are in the sample compartment.
3.
Close the sample compartment cover.
4.
Switch on the power switch.
5.
Wait for approximately 3 minutes until all
initialization is complete.
6.
Ensure that PC is correctly connected.
7.
Start UV WinLab.
ü ANALYSIS OF DATA
1.
Oncetheinstrument ison,thecomputer
maybepluggedinandturnedon.
2. Allowtheinstrument towarmupforat least 20 minutesbeforeuse.
3. On the desktop screen the PerkinElmer UV WinLab shortcut has
to be double clicked to begin.
4. When software is
openedinstrument, method has to be selected.
After selecting the instrument scan method,
DataCollection” has to be clicked under the folder list on the left-hand side
of the screen. This screenallows the user to note and change several data
collection parameters such as the data interval andwavelength range as well as
wavelengths at which there is a lamp or detector change.
5.
Click
on the sample info icon. This is where
the user will input the number of samples wish need to be run.
6. When ready to
continue, select the Auto zero button at the top of the screen. The user will
beprompted to remove samples to perform a 100%T or 0A correction. Remove any
samples fromtheinstrument, close the lidand press OK
7.
Once
complete the user will be prompted to insert the samples. The Lambda
spectrometer is a double pass instrument and has a cuvette holder for both a
blank and a sample. Place the blank in the holder nearest to the center of the
instrument and the sample in the holder nearest to the front of the instrument.
Close lid and hit OK to acquire a scan.When all samples have been run results
can be saved.
ü SHUT DOWN
1.
Close
down UV Win Lab
2.
Open
the sample compartment cover and remove samples and cells from the sample
compartment.
3.
Close
the sample compartment cover and switch off the spectrometer.
Ø MAINTENANCE
1.
The instrument is constructed with
high quality components and requires little maintenance other than to keep it
clean and free of dust. To protect the optical system from dust and fumes the
sample compartment should be kept closed when not in use.
2.
Cells should only be held by non-
optical surfaces
3.
Cells should be protected from scratches
and should never be rubbed against each other or any hard surfaces.
4.
Avoid abrasive, corrosive or
stain-producing cleaning agents, and make certain that the exposed surfaces of cells are
optically clean.
5.
Always wipe the optical surfaces of
cells dry and free of fingermarks, using a soft cloth or cleaning tissue, just
before placing them in the cell holder.
6.
A constant temperature between 15 ºC
and 35 ºC and constant humidity between 20% and 80% relative humidity need to
be maintained for proper maintenance of the instrument.
7.
Keep out of direct sunlight.
8.
Illumination with diffuse lighting
is ideal.
9.
A suitable source of electrical
power should be located in the vicinity of the instrument.
10. Electrical power must be available at a proper
earth-grounded 3-wire electrical outlet
Perkin Elmer Lamda 45
UV/VIS Spectrometer SOP
The usage of the machine is NOT
allowed without a proper training by the operator of the facility.
Ø Handling the instrument
ü START UP
1.
Open the sample compartment cover.
2.
Make sure that beam paths are free,i.e no
objects project into beam paths, no samples are in the sample compartment.
3.
Close the sample compartment cover.
4.
Switch on the power switch.
5.
Wait for approximately 3 minutes until all
initialization is complete.
6.
Ensure that PC is correctly connected.
7.
Start UV WinLab.
ü ANALYSIS OF DATA
1.
Oncetheinstrument ison,thecomputer
maybepluggedinandturnedon.
2. Allowtheinstrument towarmupforat least 20 minutesbeforeuse.
3. On the desktop screen the PerkinElmer UV WinLab shortcut has
to be double clicked to begin.
4. When software is openedinstrument, method has to be selected.
After selecting the instrument scan method , Data Collection” has to be clicked under the folder list on
the left hand side of the screen. This screenallows the user to note and change several data collection
parameters such as the data interval andwavelength range as well as wavelengths
at which there is a lamp or detector change.
5.
Click on the sample info icon . This is where the user will input the number
of samples wish need to be run.
6.
When ready to continue, select the
Auto zero button at the top of the screen. The user will beprompted to remove
samples to perform a 100%T or 0A correction. Remove any samples
fromtheinstrument, close the lidand press OK
7.
Once complete the user will be
prompted to insert the samples. The Lambda spectrometer is a double pass
instrument and has a cuvette holder for both a blank and a sample. Place the
blank in the holder nearest to the center of the instrument and the sample in
the holder nearest to the front of the instrument. Close lid and hit OK to
acquire a scan.When all samples have been run results can be saved.
ü SHUT DOWN
1.
Close down UV Win Lab
2.
Open the sample compartment cover
and remove samples and cells from the sample compartment.
3.
Close the sample compartment cover
and switch off the spectrometer.
Ø MAINTENANCE
1.
The instrument is constructed with
high quality components and requires little maintenance other than to keep it
clean and free of dust. To protect the optical system from dust and fumes the
sample compartment should be kept closed when not in use.
2.
Cells should only be held by non-
optical surfaces
3.
Cells should be protected from
scratches and should never be rubbed against each other or any hard surfaces.
4.
Avoid abrasive, corrosive or
stain-producing cleaning agents, and make certain that the exposed surfaces of cells are
optically clean.
5.
Always wipe the optical surfaces of
cells dry and free of fingermarks, using a soft cloth or cleaning tissue, just
before placing them in the cell holder.
6.
A constant temperature between 15 ºC
and 35 ºC and constant humidity between 20% and 80% relative humidity need to
be maintained for proper maintenance of the instrument.
7.
Keep out of direct sunlight.
8.
Illumination with diffuse lighting
is ideal.
9.
A suitable source of electrical
power should be located in the vicinity of the instrument.
10. Electrical power must be available at a proper
earth-grounded 3-wire electrical outlet
Perkin
Elmer Lamda 365 UV/VIS Spectrometer SOP
1.
The usage of the machine is NOT allowed without a
proper training by the operator of the facility.
Ø Handling the instrument
ü START UP
8.
Turn on power switch and allow instrument to
warm up for at least 20 minutes.
9.
Check LED light which is the lower left hand in
front of the system
Power: blue > Power on
Ready: white > The communication of between PC and system
System: white > during the checking of system status
Caution: Do not execute UV Winlab before finishing the initialization of
grating
10.
Double click UV Winlab software.
ü ANALYSIS OF DATA
8. After opening UV
Winlab , execute Scan mode icon in UVWin lab. The following message box will be
displayed
Please confirm there is empty in the
cell holder and close the sample compartment cover firmly before the
initialization.
Empty the cell holder and close the lid
firmly.Click O.K.
9.
Start
system Self Test. Check all parameters has pass click .
10. After selecting
the instrument scan method the sample
info screen will be opened. This is
where the user will input the number of samples wish need to be run.
11. After this “Data
Collection” has to be clicked under the
folder list on the left hand side of the screen. This screenallows the
user to note and change several data
collection parameters such as the data interval andwavelength range as well as
wavelengths at which there is a lamp or detector change.
12. When ready to
continue, select the Auto zero button at the top of the screen. The user will
beprompted to remove samples to perform a 100%T or 0A correction. Remove any
samples fromtheinstrument, close the lidand press OK
13. Once complete
the user will be prompted to insert the
samples. The Lambda spectrometer is a double pass instrument and has a
cuvette holder for both a blank and a sample. Place the blank in the holder
nearest to the center of the instrument and the sample in the holder nearest to
the front of the instrument. Close lid and hit OK to acquire a scan.When all
samples have been run results can be saved .
ü SHUT DOWN
4.
Close
down UV Win Lab
5.
Open
the sample compartment cover and remove samples and cells from the sample
compartment.
6.
Close
the sample compartment cover and switch off the spectrometer.
Ø MAINTENANCE
11. The instrument is constructed with high quality components
and requires little maintenance other than to keep it clean and free of dust.
To protect the optical system from dust and fumes the sample compartment should
be kept closed when not in use.
12. Cells should only be held by non- optical surfaces
13. Cells should be protected from scratches and should never be
rubbed against each other or any hard surfaces.
14. Avoid abrasive, corrosive or stain-producing cleaning
agents, and make certain that the
exposed surfaces of cells are optically clean.
15. Always wipe the optical surfaces of cells dry and free of
fingermarks, using a soft cloth or cleaning tissue, just before placing them in
the cell holder.
16. A constant temperature between 15 ºC and 35 ºC and constant
humidity between 20% and 80% relative humidity need to be maintained for proper
maintenance of the instrument.
17. Keep out of direct sunlight.
18. Illumination with diffuse lighting is ideal.
19. A suitable source of electrical power should be located in
the vicinity of the instrument.
20. Electrical power must be available at a proper
earth-grounded 3-wire electrical outlet
PERKIN
Elmer frontier FTIR Protocol/SOP
The
usage of the machine is NOT allowed without a proper training by the operator
of the facility.
·
Switch
on the power to the instrument using the switch on the rear of the instrument.
The instrument will initialize, which will take approximately 2 minutes.
·
On
the computer desktop screen double click on Perkin Elmer icon to open the
software to collect data.
·
In
the set up instrument Basic tab of the software set Start and End points of the
required scan range and the accumulatios required either as number of scans or
as a length of time. Resolution in cm-1 can also be set here.
·
In
the instrument Advance tab CO2 and H2O correction tab is
left on and acquisition is set to medium.
·
Scan
speed is adjusted to change signal to noise collection time. Slower speed
results in better signal to noise but longer collection time and faster speed
results in worse signal to noise but shorter collection time.
·
If
the results obtained are coming strange than click on restore defaults.
·
To
set up instrument data collection in auto save option save location, select
your own folder.
·
Before
collecting the data., beam path needs to be cleared for that sample chamber is
opened and it is assured that mojo is basic transmission mojo, than the lid is
closed and it is made sure that it latches. Now instrument is ready to do
background correction.
·
To
collect background, click on background icon in the software window.
·
After
background collection is done sample can be mounted.
·
To
mount a sample open the chamber place sample using double sided tape close the
lid make sure it latches.
·
Sample
name is typed in sample ID box in the software window and than scan icon is
clicked to obtain the data.
·
After
obtaining all the required data the software is closed and computer is shut
down.
v MAINTENANCE
·
Clean
the outside of the instrument using damp cloth. If necessary a mild detergent
may be used.
·
Avoid
spilling liquids into the instrument.
·
Do
not touch or attempt to clean any optical surface in the instrument because
this will impair its performance and may easily damage the component.
·
Before
moving the spectrometer, switch off the power supply wait 60 seconds and
disconnect the power cable.
·
When
all three sectors of dessicant indicator are pink than dessicant must be
changed. The instrument optics may be fogged . Do not switch the instrument
either ON or OFF until all sectors are blue.
·
Don
not use flammable gas to purge the instrument. The spectrometer contains a hot
source, and a fire and explosion will result. Only use clean dry oil free
nitrogen or air to purge the instrument.
Perkin Elmer Lamda 850
UV/VIS Spectrometer SOP
The
usage of the machine is NOT allowed without a proper training by the operator
of the facility.
Ø Handling the instrument
ü START UP
11.
. Switch on the power switch.
12.
Wait for approximately 5 minutes until all
initialization is complete.
13.
Ensure that PC is correctly connected.
14.
Place solid sample in the universal reflectance
accessory.
15.
Start UV WinLab.
ü ANALYSIS OF DATA
14. Oncetheinstrument ison,thecomputer maybepluggedinandturnedon.
15. Allowtheinstrument towarmupforat least 20 minutesbeforeuse.
16. On the desktop screen the PerkinElmer UV WinLab shortcut has
to be double clicked to begin.
17. When software is openedinstrument method has to be selected
.
After selecting the instrument scan method , Data Collection” has to be clicked under the folder list on
the left hand side of the screen. This screenallows the user to note and change several data collection
parameters such as the data interval andwavelength range as well as wavelengths
at which there is a lamp or detector change.
18. Click on the sample info icon . This is where the user will input the number
of samples wish need to be run.
19. When ready to continue, select the Auto zero button at the
top of the screen. The user will beprompted to remove samples to perform a
100%T or 0A correction. Remove any samples fromtheinstrument, close the lidand
press OK
20. Once complete the user will be prompted to insert the samples. The Lambda spectrometer is a double
pass instrument and has a cuvette holder for both a blank and a sample. Place
the blank in the holder nearest to the center of the instrument and the sample
in the holder nearest to the front of the instrument. Close lid and hit OK to
acquire a scan.When all samples have been run results can be saved .
ü SHUT DOWN
7.
Close down UV Win Lab
8.
Open the sample compartment cover
and remove samples and cells from the sample compartment.
9.
Close the sample compartment cover
and switch off the spectrometer.
Ø MAINTENANCE
21. The instrument is constructed with high quality components
and requires little maintenance other than to keep it clean and free of dust.
To protect the optical system from dust and fumes the sample compartment should
be kept closed when not in use.
22. Cells should only be held by non- optical surfaces
23. Cells should be protected from scratches and should never be
rubbed against each other or any hard surfaces.
24. Avoid abrasive, corrosive or stain-producing cleaning
agents, and make certain that the
exposed surfaces of cells are optically clean.
25. Always wipe the optical surfaces of cells dry and free of
fingermarks, using a soft cloth or cleaning tissue, just before placing them in
the cell holder.
26. A constant temperature between 15 ºC and 35 ºC and constant
humidity between 20% and 80% relative humidity need to be maintained for proper
maintenance of the instrument.
27. Keep out of direct sunlight.
28. Illumination with diffuse lighting is ideal.
29. A suitable source of electrical power should be located in
the vicinity of the instrument.
30. Electrical power must be available at a proper
earth-grounded 3-wire electrical outlet
EPR
Protocol/SOP
The usage of
the machine is NOT allowed without a proper training by the operator of
the facility.
Handling the Instrument:
Startup
(a)
Turn
the spectrometer power switch on.
(b)
Turn
the host computer on.
(c)
Start
Windows NT. Login by pressing Ctrl+Alt+Del
(d)
Enter
ID = xxx, Password =xxx and Double click on the ESR icon.
(e)
The
spectrometer control program is automatically started.
(f)
Verify
the indicators
(i) Water
(ii)
Magnet
(iii)
Gunn
(g)
ESR
standby
1)
Measurement of
the sample
(a)
Start
the instrument
(b)
The
SPECTROMETER window is diplayed.
(c)
Click
on the Q-DIP button
(d)
The
Q-DIP and SHF parameters windows is displayed.
(e)
Insert
the sample in the cavity resonator.
(f)
Set
the dial of the manganese marrker, if you will use the ESR marker
(g)
Select
the operation method, and adjust the microwave parameters.
(i) Manual operation
(ii) Automatic tuning
(iii)Semi-automatic tuing
(h)
Perform
measurement
(i)
Create
a file and save it.
(j)
Start
the data processing program, if you want to analyze data.
(k)
Shut
down the instrument
Precautions
·
When
you walk around the instrument, be aware of the cables, hoses or protruding
objects to prevent tripping, damaging or disturbing the instrument.
·
Do
not step on the frame or table of the instrument during daily operations or
maintenance. Use a steady footstool to prevent falling or damaging the
instrument.
·
Be
sure to wear protective gloves when handling a sample tube containing a
poisonous sample to avoid contact with the sample.
·
Do
not pack a sample which might explode or ignite by temperature, impact or
pressure into the sample tube to prevent the risk of injury receiving from the
shattered fragments of the sample tube..
·
Beware
of water leak i.e. after turning on the cooling water system, make sure that
there is no water leak.
·
Be
careful no to exceed the specified pressure and flow rate of cooling water,
otherwise, the cooling water tube might burst with pressure.
·
Do
not bring any iron objects near the electromagnetic field during the operation.
·
If
you wear a medical device such as cardiac pacemaker, stay out of areas where
the magnetic field is strong enough to affect it.
·
Do
not bring a magnetic tape or floppy disk near the magnet.
·
While
working in the EPR Laboratory, researchers are always required to wear Personal
Protective Equipment (PPE). The appropriate PPE include safety glasses, long
pants/skirt covering the legs completely, and closed-toe shoes.
·
Food
and beverages cannot enter the lab. Never eat or drink inside the lab.
· Samples should
be prepared in the user’s chemistry lab. The outside of sample tubes should be
cleaned with an appropriate solvent in the user’s lab.
GC-MS
Protocol/SOP
The
usage of the machine is NOT allowed without a proper training by the operator
of the facility.
Handling of Instrument
·
Make
sure that carrier gas (Helium or Nitrogen gas) cylinder is open and flowing at
7.0 bar.
·
Turn
on the main power switch of the GC, MS, PC and printer.
·
Double
Click 'GCMS' Real time analysis' icon from the main screen.
·
Click
'Instrument from the main menu and select 'vacuum control".
·
Click
"Auto startup' in the 'vacuum Control window and wait until the
function is completed.
· Open the Method
file' required for the analysis and click download initial parameters from 'Acquisition'
in the main Manu.
·
It
display in green signal as 'GC Ready & MS Ready.
·
Open
Peak View monitor screen from assistant bar.
·
Click
Start Auto Tuning' from the Assistant bar, once tuning finished save the Tuning
file from 'save as with today date then click 'Auto tuning Report.
·
Click
'sample login from Assistant Bar, and then enter the sample name, datafile,
analysis details etc. and current Tuning File' then click OK.
·
Click
'Stand by' from the Assistant bar and wait till GC & MS Ready display on
the screen.
·
Inject
1 μl sample and press Start button from GC control panel.
Note: (a) The pressure value should reach upto 2.5
x 10-5.
(b) Sample must
be volatile.
(c) Boiling
point must be <300oC.
(d) After
analysis the temperature should go down to fix 100.
2.
Analysis of Data
·
After
completion of analysis. Click GCMS post run analysis' in Desktop.
·
Go
to File' in the main menu and open the Data required for the report preparation
and do the integration.
·
Click
peak table in the side menu and click Select all and register in the Edit menu,
·
Tab
to the next window and click Similarity search.
·
Click
Save from the main menu and close peak table.
3.
Shutdown:
·
Click
"Instrument' from main menu and select "Vacuum
control" icon.
·
Click
'Auto shutdown' in the window, wait till Auto shutdown 'completed appear
on the screen.
·
Close
'GCMS Real Time Analysis' window and shutdown the PC.
·
Turn
off the MS, GC, PC and printer
·
Close
the 'Cylinder' tightly.
4.
Precautions:
·
Temperature
should be maintained between 15°C and 27°C respectively.
·
Humidity
should be maintained between 40 % and 60 %.
·
Purity
of helium or nitrogen used for analysis should be A grade.
·
Change
the vacuum pump oil for every 4-6 months.
·
Replace
the injection septa for minimum of 100 injections.
·
Keep
the split/spit less liners clean and free from deposition.
·
Cleaning
is recommended when the deposition is high and reduced response, peak
tailing/mounting/fronting is observed.
·
Monitor
the leak parameters for air and water every day.
·
Food
and beverages cannot enter the lab. Never eat or drink inside the lab.
·
Samples
should be prepared in the user’s chemistry lab. The outside of sample tubes
should be cleaned with an appropriate solvent in the user’s lab.
·
Be
sure to wear protective gloves when handling a sample tube containing a
poisonous sample to avoid contact with the sample.
Powder
X-ray Diffraction Protocol/SOP
1. The usage of the
machine is NOT allowed without a proper training by the manager of the
facility.
2.
General introduction
a) Figure 1 shows
the XRD main unit. The X-ray warning lamp (#4) will light if the equipment is
energized. To open the door, press the door-lock button, wait for 5 seconds,
then slide the door open. Click the button again when the door is closed.
b) There is an
indication light on the X-ray tube as shown in Figure 2. When it is on, it
means the shutter is open and that an experiment is in progress. Please refer
to the online scheduler in order to reserve a time to use the XRD.
3.
Starting the XRD
(You may skip step a. c. d if the system
and X-ray is already on. When the X-ray light on top the system is lit, it
means the X-ray was left on, you don’t need to turn it on again.)
a)To turn on the Main unit, press
the “|” ON button from the main panel as shown in Figure 1. Then the OPERATE
light should light up in green.
Figure. Main panel
b) The computer on
the desk to the right of the XRD controls the SmartLab software. This computer
should always be left on.
c) Turn on the
Haskris water chiller located to the left of the XRD unit by flipping the
ON/Off button. Wait for at least 30 minutes before turning on X-ray Tube. THIS
IS VERY IMPORTANT! DO NOT FORGET THIS STEP!
d) The X-ray tube
may be turned on after the 30 minutes wait. From the flow chart section of the
software as indicated in Figure 5, click “Startup”. In the new window that
appears (see Figure 7b) click “execute” to start the X-ray tube. There is no
need to change any parameters in this window. This process may take 5 minutes
4.
General Procedure:
- Switch on
the UPS.
- Open the
chiller and wait until the operating temperature (21oC) is reached
- Adjust the
water pressure at46psi.
- Turn on the
computer.
- Turn on the
XRD instrument by switching on the MCB.
- Prepare the
sample slide for analysis.
- Insert the
USB in computer and run the software.
- Wait for 15
minutes until the instrument is ready for analysis.
- Place the
slide between the X-ray beams and analyze the sample..
5.
Precautions:
- Follow
the
SOP strictly to keep the instrument in good condition. No explorations allowed on
software unless permitted by lab operator.
- Never
use
your own USB drive on the XPS computer.
- Never
surf
the web on the XPS computer to minimize the risk of the computer being
hacked.
- Only trained personnel are
permitted to operate XRD unit.
- Be aware of the xray beam
path and never lace any part of the body in the direct beam path.
- Food and beverages cannot
enter the lab. Never eat or drink inside the lab.
- Samples should be cleaned
with an appropriate solvent in the lab.
- Be sure to wear protective
gloves when handling a sample tube containing a poisonous sample to avoid
contact with the sample
- To insure your safety, do
not attempt any unauthorized repair of x-ray unit. ··
- Be sure the beam is off and
shutter is closed prior to sample changing or other activity. Check all
warning lights prior to placing hands near the beam line. Use a GM
radiation survey instrument to confirm “beam off” conditions. · Use the
shielding and interlocks provided. Do not bypass interlocks.
- Users should acknowledge Department of
Chemistry in their publications.
SOP FOR CIRCULAR
DICHROISM
I. Start up
1.
Turn
on Nitrogen gas flow.
2.
Turn
on the computer and open the program “Spectra Manager”
3.
Turn
the power switch on the CD to “ON” position (note: this will not turn the lamp
on)
4.
Start the “Spectra Manager Program” and double
click on “Spectrum Measurement”.
II. Data Acquisition
1.
On
the drop-down menu bar, select the “Parameters” tab. The following are only a
guideline for peptide/protein solutions.
(i)
Bandwidth = 1 nm.
(ii)
Response = 4 sec (1 sec)
(iii)
Sensitivity = Standard.
(iv)
Measurement Range = 250 – 190 nm.
(v)
Data Pitch = 0.1 nm.
(vi)
Scanning Mode = Continuous.
(vii)
Scanning Speed = 100 nm/min.
(viii)
Accumulation = 5
2.
At
the “Data Mode” tab, ensure the settings are appropriate for your experiment:
For CD experiments:
Channel #1 =
CD
Channel #2 =
HT
3.
Click “OK” to apply the parameters and close the dialog box.
4.
Click “Start” to begin data acquisition.
5.
Once a scan is complete, it will automatically open up in “Spectra Analysis”.
This software can be used to perform background subtraction and peak fitting.
III. USING the
Peltier temperature controller
1.
Turn
on the water bath and make sure that the temperature is set to 20°C.
2.
Turn
the power switch on the Peltier controller (located between the PC and CD) to
“On” position.
3.
Press
the “Start” button on the Peltier controller.
4.
In “Spectrum Measurement”, go to the menu
Measurement→ Accessory
5.
Choose “Temperature” and change the option to
“JASCO Peltier Type (Single, RTE On)”
6.
You
can now use Control → Accessory menu in the “Spectrum Measurement” software to
set the temperature.
7.
When
you are done power down the Peltier controller and the water bath.
IV. Shut down
1.
Leave
the Nitrogen gas ON.
2.
In
the “Spectrum Measurement” software go to the menu Control → Light Source.
Uncheck “Lamp On” box and click OK. This will turn the lamp off.
3.
Quit
the software and turn off the main green power switch.
4.
Now,
turn the Nitrogen gas off.
# Always
Remember:
1.
Do
not use the CD without training!
2.
Do not change settings on the regulator or the gauge
itself.
3.
Always
leave the nitrogen stream on when the instrument (specifically the instrument
lamp) is on! Once the instrument is
fully shut down, wait a few minutes and only then turn the nitrogen off.
4.
The
lamps should never have > 1000 hours of use.
5.
If
case of Peltier controller, Do not allow water to circulate when the lamp is
not lit!
Fluorescence
Spectrophotometer SOP
I. Start up
5.
Turn
on the spectrophotometer and the computer.
6.
Double-click
the “FL Solutions” program icon. A window will appear which displays all tool
bars.
II. Parameters
settings
1.
From
the R.H.S tool bar, select the “Method” command then follow the steps:
(a) General>
Measurement> Wavelength scan
(b)
Instrument> Scan mode> Emission
(c) Data
mode> Fluorescence
(d) Set
excitation and emission wavelengths, Slit widths, PMT voltage and scan speed
values and then click “OK” to apply the parameters.
2.
Select
Sample> Select the specified folder (already created in the D: file)>
write the sample name for each run>OK.
3.
Select
Monitor> Put Y-axis max and min values > Ok
III. Sample run
1.
Clean
the cuvette properly and fill it with blank solution.
2.
Keep
the cuvette in the chamber and click the “auto zero” button.
3.
Now
fill the cuvette with sample solution and click “measure” button for each run.
The instrument records the spectra within the specified wavelength range.
4.
The
spectra and data will be saved in the specified folder.
5.
Open
the data processing window after data acquisition via File > Open. A spectrum alone can be displayed
or both spectrum and peak table can be displayed.
IV. Shut down
After
completion of measurement, shut down the instrument in the following procedure
5.
Select
File > Exit command.
6.
A
window will appear from where choose the second option.
(i) Close the monitor window but keep the
lamp operating?
(ii) Close
the lamp and then close the monitor window?
7.
Click
Yes, and the FL Solutions program will then be terminated.
8.
Allow
the light source to cool for approximately five minutes.
9.
Confirm
that the sample chamber is empty and then Turn off the power switch of the
spectrophotometer main unit.
10.
Then
Turn off power to the personal computer.
# Always
Remember:
6.
Do
not use the fluorescence spectrophotometer
without proper training!
7.
Keep
the lamp off when you are not using spectrophotometer.
8.
Check
the usage period of the Xenon lamp. It should not exceed 500 hours of use.
9.
Do
not smoke or hold a flame near the fluorescence spectrophotometer.
ISOTHERMAL
TITRATION CALORIMETRY (ITC) SOP
I. Start up
7.
Turn
on the computer
8.
Turn
on the ITC instrument (there are 2 power switches, one at the back)
9.
Then
open the MicroCal iTC200 software. Verify that the red light at the
front of the instrument is on.
II. Cleaning of
cell & syringe
1. Make sure you
have wash station bottles and filled to at least 50% with the appropriate
solutions i.e with distilled water and methanol respectively.
2. Select the
Instrument Controls tab and click the Cell and Syringe Wash button in the
Washing Module panel to clean the cell and syringe.
3. Follow the
step-by-step screen instructions and then click OK to start the washing
procedure of cell as well as syringe.
4. While the
syringe washing procedure is running, rinse the sample cell manually with blank
solution a number of times.
5. Repeat these
cleaning steps after completion of experiment/titration.
III. Cell and
syringe filling
1. Fill the sample cell and reference
cell carefully with water and sample solution, respectively by the Hamilton
syringe. Be careful not to introduce air when doing this.
2. Click on the Syringe Fill button in the Washing module panel and follow the
step-by-step instructions on screen to load the syringe with ligand solution.
3. Insert the pipette into the cell
port. Apply light pressure on the top to make sure it is correctly seated.
Click OK.
IV. Experimental
set up
1. Click on the
“Advanced Experimental Design” tab.
2.
Set the Experimental Parameters as follows and set the Feedback Mode/Gain to
High.
3.
Set the Edit Mode to All Same and enter the following injection parameters
4.
Enter a file name for the results (e.g. Training.itc) in the Data File Name
field.
5.
After filling cell and syringe, click Start button to begin the titration. The
run time will be approximately 1 hour.
6. Click on the
Real Time Plot tab to view the raw data.
V. Data analysis
1.
Double
click on the MicroCal Analysis Launcher icon on the desktop.
2.
Click
on iTC200. This will open the analysis software for Microcal iTC200.
3.
Click
on Read Data. To open your titration experiment data file Training.itc, click
on the data file, click on Add File(s) and then on OK.
4.
Click
the model fitting to fit the data. This will open a dialogue box showing the
initial fitting and values for stoichiometry (N), association equilibrium
constant (K) and enthalpy change (H).
5.
The
N, Ka, ΔH and ΔS values based on the fitting will be displayed next
to the plot.
6.
Select
File>Save Project As to save the fitted results. Enter a name and click
Save.
VI. Shut down
1.
Close
the software.
2.
Shutdown
Personal Computer.
3.
Turn
off 2 switches of ITC.
Always Remember:
1.Do not work with the device unless you have proper knowledge of handling.
2.The system should be cleaned after each run using the Cell and Syringe
Wash command.
3.Replace the distilled water in the reference cell every week.
4.There should be NO air bubbles inside the cell during the experiment run.
5.do not invert the Hamilton syringe.
ATOMIC ABSORPTION SPECTROMETER, GBC-932-PLUS,
AUSTRALIA
Objective
This document outlines the procedure for the operation of the
Atomic Absorption Spectrophotometer from GBC Company (Model 932-Plus),
Australia.
Scope
Atomic absorption spectrometry (AAS) is a technique in which free
gaseous atoms absorb electromagnetic radiation at a specific wavelength to
produce a measurable signal. it is an important sensitive method that is
suitable for the determination of selected elements at the trace and partly at
the ultra-trace level.
Operating
procedure
Ø Turn on the instrument (Power switch at the right hand of the
instrument is only turned on when we wish to use furnace), computer and verify
ventilation.
Ø Double click on the “GBC Avanta” program located on the desktop.
Ø From the drop-down menu select the “online” item and press OK. The
program will proceed through a checking process. Please wait until finishing
the checking process. This usually takes a few minutes.
Ø An element selection window will appear. Select the working lamp
and click on finish.
Ø GBC Avanta” software will open. Open the “instrument” menu and
select the “measure method” item. The “set measurement method” dialog box
opens. Select the method you wish to use then click on Execute. Wait for the
system to locate the selected atomizer in the path of light. This usually takes
a few minutes. After finishing this procedure, close the “set measurement
method” dialog box
Ø Optimize atomizer position using instrument” menu-Burner
parameters. The position of the atomizer should be adjusted carefully so that
the whole energy of the hollow cathode lamp reaches the detector. The path of
light can be checked by a piece of white paper.
Ø After adjusting the position of the atomizer, click on “Lamp” icon
and then click Next. Press “search peak” button. The “wavelength scanning/peak
searching” dialog box opens.
Ø Click “search peak” button and wait. This usually takes a few
minutes. After appearing the spectrum, close the “wavelength scanning/peak
searching” dialog box.
Ø Click on “Energy” icon. The “Energy” dialog box opens. Click on
“Auto-balance”. Energy of the lamp should be about 100 %.
Ø Click on “Sample” icon and enter calibration and sample
information.
Ø Prepare standard solution (dilute) for calibration. Blank should be
prepared for adjusting zero (0)
Ø Run the prepared sample.
Do’s and Don’t
Ø Do not leave without closing the acetylene tank when not in use.
Acetylene is explosive and quite dangerous.
Ø To shut off the flame in an emergency- push the red button on the
right-front of the instrument.
Ø The pressure of the acetylene gas cylinder must never fall below 4
bar.
Ø When the flame is ignited, to protect the furnace, insert the steel
plate which separates the furnace from the flame.
Ø Water trap should be filled with water, so that excess amounts of
water overflow into the waste container.
THERMOGRAVIMETRIC
ANALYSIS/DIFFERENTIAL SCANNING
CALORIMETRY, SHIMADZU, JAPAN
Purpose
and Scope:
This document describes the procedures
and for using the TG/DTA-60 (SHIMADZU C30574300134)/DSC-60 (SHIMADZU
C30454400547) series Instruments TGA/DSC. The scope of this document is to
establish user procedures. TGA stands for ‘Thermogravimetric Analysis’. A very
versatile and sensitive instrument to measure the change in weight as a
function of temperature. From TGA/DSC we can determine the number of dynamic
properties. DSC stands f6r differential scanning calorimeter used to study
various endothermic and exothermic reactions in the sample as well as thermal
transitions.
Ø Thermal
stability of materials
Ø Oxidative
stability of materials
Ø Estimated
lifetime of a product
Ø Decomposition
kinetics of materials
Ø The
effect of reactive or corrosive atmospheres on materials
Ø Moisture
and volatiles content of materials.
Ø Thermal
transition
Ø Exothermic
and endothermic reaction
These properties can be determined from
either weight gain or weight loss. It is also important to remember that this
is a kinetic measurement so rates and masses should be carefully chosen and
they should be included in any reporting of data.
Features and Specifications of the
Instruments DTG-60 (SHIMADZU C30574300134):
Ø RT-1000
C
Ø N2
and/or dry air purge
Ø Mass
accuracy: +/- 0.5%
Ø Mass
precision: +/- 0/01%
Ø Sensitivity:
0.1 ug
Do’s
and Don’t
Ø Please
do not eat or drink while operating this instrument.
Ø Never
touch the hang-down wire or thermocouple.
Ø Never
attempt to operate the instrument without nitrogen flow.
Ø Do
not change any of the user preferences or instrument preferences.
Ø Notify
the staff member of any issues; do not attempt to ‘fix a problem.
Ø Once
a measurement is running it is very important not to bump or disturb the
instrument or the table it’s sitting on, it is very sensitive to vibrations and
can cause erroneous or anomalous behaviour.
Ø Before
starting, please verify that there is plenty of N2 in the cylinder
and that the flow rate is at 10/90 mL/min for the furnace.
Procedure steps
Ø Turn
on the stabilizer and computer
Ø Switch
on the instrument, and wait for 10 mins.
Ø Calibrate
the detector to a zero (0) position
Ø Left
the instrument on for about 20 mins. After calibration.
Ø Again,
calibrate the detector to zero (0) position by placing alumina pan on it.
Ø Place
the sample (4-8 mg) in the right alumina pan.
Ø Adjust
the operating temperature, suggested by the scholar
Ø Set
the parameter
Ø Make
a folder by name of supervisor and scholar
Ø Code
the sample
Ø Run
the instruments for the analysis
Ø Use
the nitrogen gas for purging during the analysis
ELECTROCHEMICAL
ANALYSIS SYSTEM, VersaSTAT 3, AMETEK SCIENTIFIC, USA
Cyclic Voltammetry is a versatile method for scientific
investigation and innovation due to the fact that most processes involve
electron transfer, which makes them be able to be monitored by this technique.
Beginning Experiments:
Ø Turn on the VersaSTAT 3 Instrument Electrochemical Workstation. The
switch is a button on the front, facing the small box, and the button will be
green when it’s on. Make sure to turn it off at the end of use.
Ø Open the VersaStudio software located on the desktop of the
computer.
Ø Click Setup
and Techniques to choose your method of electrochemistry, e.g. cyclic
voltammetry (CV)/Chronoamprometery/Chronopotentiometry/LSV etc.
Ø
Click Parameters to set up your method, and save the
change.
·
For CV, set initial and final E
(usually 0.00V), and high and low E will be determined by the desired potential
window for your experiment. Scan polarization will depend on which direction
you want to scan first, negative or positive. The scan rate is usually set as
0.1V/s, and you can change this to the desired scan rate based on your needs.
Set the sensitivity to be the value bigger than what you think the current will
be. It can stay as default, 10-6, or you can change it to a lower
sensitivity setting, preferably 10-5. See the VersaSTAT 3 point-person for questions. Sweep
segments are usually set as 5. Sample interval and quiet time will stay
asdefault.
Ø
Once the parameters are set, click the Play button to
start the measurement. You
can Pause and Stop a run.
Ø
When the measurement is done, make sure to save the
data. It will say Run
Unsaved at right if you haven’t saved.
Ø Data save as
a VersaSTAT 3 Instruments
file, you can download the software if you have aPC and analyze data with their
program or you can convert the data to text.
Ø If
the data is very noisy, smoothing can be performed by the software by clicking
DataProc
and smooth.
· The noise of the data can also be due to the fluctuation of the glovebox atmosphere. If this is the case, repeated measurement is recommended.
Do’s and
Don’t
Ø Do not touch
the electrode during the analysis.
Ø There should
be no electricity fluctuations.
Ø Only see a
flat line without signals. Check whether your electrodes are connected
properly.
Ø Only see
noise or weird curves-check whether electrodes are connected properly and are not
touching the side or each other, and are submerged down into the solution.
Ø When you open
the program you get an error message-shut down the program and turn on VersaSTAT 3workstation first, then open the
program.
Ø If you see
the signal at first and then it flats out-stop the run and set your sensitivity
to a
higher value, or lower value; you are just out of the range
of the current you chose.
After
Experiments:
Ø Used Pt wire counter electrodes need to be soaked in 1M HNO3
for at least 1 h.
Ø Used Ag wire reference electrodes with the glass tubes need to be
rinsed by the solvent used for Echem data collection at least 3 times and
soaked in 1M HCl for about 1 h.
Ø Used working electrodes (Pt or glassy carbon or graphite) need to
be rinsed by the solvent used for Echem measurement at least 3 times.
Ø Used Ag/AgCl reference electrode will be soaked in the solvent used
for experiments, DMF, and MeOH for 5 mins respectively. After that, rinse the
electrode in distilled water and put it back in a vial with fresh 1M KCl
solution.
Ø Turn off the VersaSTAT 3 Instrument Electrochemical Workstation
after use.
MAGNETIC SUSCEPTIBILITYBALANCES SHERWOOD
SCIENTIFIC, UK
Identification
of the method
Operation of
the Magnetic Susceptibility Meter (SHERWOOD).
Applicable
matrix of the method
This instrument
can be used for soil/sediment samples.
Scope and
Summary of the Method
This standard
operating procedure is required to operate the SHERWOOD Magnetic Susceptibility
Meter. This method is used to determine the magnetic susceptibility of soil and
sediment collected from the field or artificial mixtures of mineral or chemical
ingredients.
The SHERWOOD
Magnetic Susceptibility System comprises a portable measuring instrument, the SHERWOOD
meter, and three sensors controlled by a handheld Trimble Nomad field data
logger loaded with Bartsoft control software. A low-frequency AC magnetic field
is generated by the SHERWOOD Magnetic Susceptibility
System. When a sample is placed in this field, there is a change in the field
that is detected by the system. This change is converted into a magnetic
susceptibility reading. Use SI units when taking measurements.
Detection Limit
The SHERWOODMagnetic
Susceptibility Meter is capable of measuring the magnetic susceptibility System
International Units (SI). The maximum resolution is 2x10-6 SI (vol)
depending on the attached sensor and environment. The operating temperature
range for this instrument is -40° C to 70° C.
Procedure Steps
Ø Turn on the instrument.
Ø Wait until the instrument is calibrated.
Ø Weigh the empty capillary.
Ø Weigh the capillary with the sample in it more than 1.5 cm.
Ø Run the instrument and note the reading.
Do’s and Don’t
1.
Temperature-induced
drift - Each sensor compensates to minimize temperature-induced drift that
arises from changes in the permeability of free space which affects the
calculation for magnetic susceptibility.
2.
Wet conditions
- Very wet conditions should be avoided, but the instruments are sealed to
prevent the entrance of moisture.
3.
This instrument
should not be operated close to high power radio transmitters or heavy
electrical machinery.
4.
If measurements
fluctuate, the removal of the transit clamp may alter the zero setting of the
beam.
5.
Should it not
be possible to zero the display using the zero knob.
6.
The balance
should be set up on a flat, stable surface free from vibration, and should be
positioned away from the influence of stray magnetic fields.
7. The sample guide is designed such that a glass tube that is broken
while using the balance will fall through the base of the unit. Care should be
taken not to damage the thin wall of the guide tube when cleaning.
DIGITAL AUTOMATIC POLARIMETER (M-HB-P701) HELIX
BIOSCIENCES, NEW DELHI, INDIA
Objective:
This SOP gives the procedure of operation and calibration of polarimeter.
Scope:
The operation and calibration of Polarimeter instrument to determine the
specific optical rotation of given test sample.
Operating
Procedure
1.
Switch On the electric supply and within 10
minutes, the sodium lamp to glow optimally.
2.
Check the polarimeter tube and its cabinet. It
should be clean.
3.
Position the incoming light source, polarimeter
tube and cabinet in a line by rotating the revolving stand of sodium lamp.
4.
For clear vision of colour shade, adjust the
upper and lower eyepiece to anterior and posterior side.
5.
Fill the polarimeter tube with water / blank in
which the substance is dissolved and place this in cabinet. (Determine at
Temperature 25°C, unless & otherwise specified in the procedure/monograph)
6.
Rotate the knob at the anterior right and match
the colour shade in both sections by
observing
in lower eyepiece.
7.
Take the reading in upper eyepiece, it should
be about zero. Rotate the vernier scale
at the
left anterior head of instrument. Add the reading of main scale and vernier
scale,
and find
the reading of rotation.
8.
Fill the polarimeter tube with the liquid /
prescribed solution. Solution / liquid should be clear and free from
air-bubble.
9.
Rotate the knob for matching the colour shade
similarly as step 6
10. Take the reading in + or – in upper eyepiece
and note it. (Take three concurrent readings of blank as well as sample and
consider the mean value).
Do’s and Don’t
Ø Handle the instrument and polarimeter tube carefully.
Ø Do not keep polar meter’s tube without cleaning after use. It
should be clean after each use.
Ø Keep Switch off the instrument when not in use.
Ø The solution should be homogenized.
Ø Specific optical rotation to be measure sample should be clear.
Ø Avoid air bubbles in sample tube, and if any small air bubble
observed, then should be adjusted in the middle portion of the tube.
Ø The cell should be placed properly.
Ø Read the scale properly.
Ø The cell should be clean properly.
Ø The polarimeter should be placed in a flat surface.